This advanced analyzer integrates MOCON’s proven WVTR analyzer system with a precision environmental chamber to elevate package testing to the next level in accuracy, repeatability and ease of use. With special functions designed for package testing they greatly reduce overall test time while the new package testing cartridge designs offer simple sample preparation and improved repeatability of test results. They feature 4 independent cells with changeable test cartridges that allow a wide variety of package sizes and types to be easily and accurately tested, many without the use of epoxy.

Request a Quote

The AQUATRAN 3/40 package testing analyser can test the Water Vapor Transmission Rate (WVTR) on both films and packages from low to high barriers. It uses the MOCON’s patented long life modulated IR sensor which is the basis for the ASTM standard F1249.  The 4 independent test cells can perform different tests in parallel under precisely controlled environmental conditions from 10 to 50 degrees C and a range of 30 to 90% humidity in the chamber.

A variety of testing cartridges are available to easily and accurately test trays, bottles, flexible containers, pouches, cups, pods, caps, closures, cartons, films and just about any other type of sample. Custom designed cartridges are also available to simplify sample preparation and improve testing repeatability.

For more than 25 years, MOCON’s water vapor series of instruments have been the most trusted and longest lasting on the market. The AQUATRAN 3/40 is the newest option available that takes whole package testing to the next level allowing brand owners and packaging service providers the ability to reliably test the shelf life of their finished packages. It is the perfect complement to any testing laboratory who frequently tests complete packages.

Highest Accuracy Package Testing

  • MOCON’s Patented Modulated Infrared sensor is the only modulated sensor system on market meeting ASTM F1249
  • Low detection levels down to 0.000025 g/(pkg x day)
  • Fully automated temperature, flow and RH controls simplify testing and provide unparalleled repeatability

Ease of Use

  • Removable test cartridges allow fast sample exchange with the touch of 1 button
  • Different cartridge options designed to make mounting and testing for a wide variety of package types simple, accurate, and repeatable
  • Advanced automatic purge and conditioning capability for packages reduce overall testing time
  • Touch screen interface allows complete control and monitoring of all testing variables throughout the testing process

Variety of Package and Film Test Cartridges

  • Easy sample prep on the benchtop, many without epoxy
  • One button quick release and loading
  • Single use mounting foil plates simplifies package mounting
  • Specialized cartridge options save time and improve test repeatability

Automated Control Features

  • Flow, Temp, & RH control
  • Test Process Development and Recommendations
  • Programmable Test Sequencing
  • Test report printing and export
  • Sensor Protection

Touchscreen Interface

  • Intuitive automated operation requires less training
  • Built in multi-level user management and data security
  • Event log & trouble shooting guide
  • Save on valuable lab space by eliminating need for PC

Modulated IR Sensor

  • Short exam times reduces overall test time to increase throughput
  • Longer sensor life than others on the market
  • Accurate and repeatable results
  • Most trusted modulated IR sensor that fully complies with ASTM F1249

Advanced Package Testing Features

  • Automatic quick purge to desired settings
  • Automatic high flow conditioning mode

TruSeal®

  • Eliminate Individual Zero
  • Reduced testing time (increase throughput)
  • More repeatable results
  • Allows tool-less mounting of cartridges

Warranty

  • Full 100% parts and workmanship warranty for 12 months
  • Exclusive 4-year warranty on the Modulated IR sensor

Capabilities – Model Specific

Units 3/40 L
Test Ranges
g/(m2 x day) 0.005 to 100
g/(pkg x day) 0.000025 to 0.5
Resolution g/(pkg x day) 0.000001
Repeatability g/(m2 x day) 0.005 or 2%*

*whichever is greater

Capabilities – All Models

AQUATRAN 3/40 Units
Test Temperature Range 10° C to 50° C ± 0.2 ° C
Controlled RH Testing Range 30 to 90% ± 5%
Chamber Size 25.5″W x 17″D x 12″H

* See RH and Temperature Range chart in brochure for details

Other Specifications:

Physical Requirements

  • 35.5” H x 33.0” W x 29.75” D (88.9 cm x 83.8 cm x 75.6 cm)
  • 225 Lbs. (102 kg)

Electrical Requirements

  • Voltage: 100 – 240 VAC 50/60 Hz
  • Maximum Power Draw: 500 VA

Laboratory Requirements

Temperature Operation 22 °C ± 2 °C
Storage 10 °C to 30 °C
Humidity Operation 10% to 60% RH (non-condensing)
Storage 5% to 85% RH (non-condensing)

Gas Supply Requirements

Carrier Gas Gas Composition Nitrogen (99.7% or better)
Supply Pressure, Nominal 25 PSI, (1.72 Bar), (172 kPa)
 Test Gas Gas Composition  Dry compressed Air (ISO 8753-1:2010[1:3:1])
Supply Pressure, Nominal  85 PSI, (5.86 Bar), (586 kPa)

Connectivity

  • Remote Access and Control (requires optional Perm-Net Light software)
  • Standard Ethernet port (100T)
  • Dual USB 2.0 ports (for connection of keyboard, mouse or memory devices)

Compliance Standards

This product is UL and CSA listed using the ETL mark from Intertek Group plc. and is certified to conform to CE.

Conforms to the following internationally recognized safety standards.

  • IEC 61010-1:2010 Third Edition
  • UL 61010-1:2012 Third Edition
  • CAN/CSA-C22.2 No. 61010-1:2012 Third Edition

Conforms to the following internationally recognized EMC standards.

  • EN 61326-1:2013
  • EN 61000-3-2:2006, A1:2009, A2:2009
  • EN 61000-3-3:2013

Conforms to EU Directive 2011/65/EU RoHS Compliance.

Cartridges & Options

  • 80 cm2 package testing cartridge
  • Large vertical tray testing cartridge
  • Flexible pouch testing cartridge
  • 50 cm2 film testing cartridge
  • Package adaptor cartridge
  • Conditioning rack inside chamber
  • Custom engineered cartridges

Consumables

  • Cert Films 1-4 & set
  • Single use package mounting foil plates

Software 

  • 21 CFR Part 11 user and data security
  • Remote viewing software from a PC – PermNet Lite
  • PermWare Permeation Laboratory Management Software

Service & Support Options

  • IQ/OQ documentation & validation
  • Basic Preventative Maintenance Plan
  • Plus Preventative Maintenance Plan

Package Films

  • Monolayer
  • Multi-layer
  • Coated Films
  • Coated Paper/paperboard

Packages

  • Common food packages
  • Medical packages
  • Bottles/Closures

Test Standards Compliance

ASTM F1249
TAPPI T557
JIS K7129
ISO 15106-2
GB/T 26253-2010
GB/T 31355-2014

Test Methodology

Following ASTM F1249

Standard Test Method for Water Vapor Transmission Rate Through Plastic Film and Sheeting Using a Modulated Infrared Sensor

Summary of Test Method
1. A dry chamber is separated from a wet chamber of known temperature and humidity by the barrier material to be tested. The dry chamber and the wet chamber make up a diffusion cell in which the test film is sealed. Water vapor diffusing through the film mixes with the gas in the dry chamber and is carried to a pressure-modulated infrared sensor. This sensor measures the fraction of infrared energy absorbed by the water vapor and produces an electrical signal, the amplitude of which is proportional to water vapor concentration. The amplitude of the electrical signal produced by the test film is then compared to the signal produced by measurement of a calibration film of known water vapor transmission rate. This information is then used to calculate the rate at which moisture is transmitted through the material being tested.