PERMATRAN-C ® 4/30
Carbon Dioxide permeation rate measurement system for films and packages.
The PERMATRAN-C 4/30 CO2 permeation analyzer can test the Carbon Dioxide Transmission Rate (CO2TR) on both films and packages from low to high barriers. It features two independent cells with interchangeable horizontal test cartridges that allow for a wide variety of sample sizes and packaging materials to be easily and accurately tested.
The PERMATRAN-C 4/30 CO2 permeation analyser can test the Carbon Dioxide Transmission Rate (CO2TR) on both films and packages from low to high barriers. It features two independent cells with interchangeable horizontal test cartridges that allow for a wide variety of sample sizes and packaging materials to be easily and accurately tested.
High Accuracy
- MOCON’s patented modulated infrared sensor is the only modulated sensor system on the market meeting ASTM F2476
- Low CO2TR detection levels down to 0.5 cc/(m2 x day)
- Fully automated temperature, flow controls simplify testing and provide unparalleled repeatability
High Throughput
- Modulated IR sensors measure CO2 gas concentrations in real time greatly reducing overall CO2TR test times compared to Zahm-Nagel CO2 testing methods
- Excellent correlation to Zahm-Nagel shelf life studies allowing equivalency while significantly reducing long term testing
- Automated test methods can start a test by a single press of a button
- TruSeal gas flushed perimeter sealing can eliminate the need for absolute zero in many tests, thus shortening overall test times
The PERMATRAN-C 4/30 analysers can test the Carbon Dioxide Transmission Rate (CO2TR) from low to high barriers on films, packages and bottles. Unlike manometric or Zahm-Nagel methods that require long periods of time to provide results the PERMATRAN-C 4/30 utilizes MOCON’s patented modulated IR sensor system allowing it to provide accurate results in much less time.
Designed for a wide range of barriers, this analyser provides highly repeatable test results from 0.5 to 8,000 cc/ (m2 day). The upper detection limit can be extended with the use of a reduced area film test cartridge (factor x9), or by using various sizes of masking.
The analyser has industry leading features such as removable and interchangeable horizontal test cartridges, easy to use touch screen operator interface, and fully automated testing providing higher throughput in the lab with less training.
For more than 25 years, MOCON’s PERMATRAN series of analysers have been the most trusted and longest lasting on the market. The 4/30 delivers fast CO2TR results in less time than traditional methods helping to speed up your research and development of new materials or give you efficient QC verification of existing products.
Removable Cartridges
- Easy sample preparation on the benchtop
- More consistent film loading into the analyser
- One button release and loading
- Pneumatic clamping provides more repeatable results
- Interchangeable cartridge options that expand the testing capability
TruSeal®
- Eliminates Individual Zero
- Reduces testing time (increases throughput)
- More repeatable results
Automated Control Features
- Flow, Temp control
- Test Process Development and Recommendations
- Programmable Test Sequencing
- Test report printing and export
- Sensor Protection
Touchscreen Interface
- Intuitive automated operation requires less training
- Built in multi-level user management and data security
- Event log & trouble shooting guide
- Save on valuable lab space by eliminating need for PC
Modulated IR Sensor
- Short exam times reduces overall test time to increase throughput
- Long sensor life
- Accurate and repeatable results
- Most trusted modulated IR sensor that fully complies with ASTM F2476
Warranty
- Full 100% parts and workmanship warranty for 12 months
- Exclusive 4-year warranty on the Modulated IR sensor
Capabilities – Model Specific
Units | Model 4/30 L | |
Test Ranges (50 cm2) | cc/(m2 × day) | 0.5 to 8,000 |
cc/(100 in2 × day) | 0.03 to 516 | |
Test Ranges (5.6 cm2) | cc/(m2 × day) | 4.5 to 71,000 |
cc/(100 in2 × day) | 0.29 to 4,600 | |
Test Ranges (5.0 cm2 masked) | cc/(m2 × day) | 5 to 80,000 |
Resolution (50cm2) | cc/(m2 × day) | 0.0001 |
Test Range – Package | cc/(pkg × day) | 0.0025 to 40 |
Repeatability | cc/(m2 × day) | 0.25 or 2.0%* |
Temperature Control Range | °C | 10 to 40 ± 0.2 |
Humidity (RH) Control Range Film Cell, Test Gas |
% RH | Dry Only |
*Whichever is larger
Capabilities
PERMATRAN-C 4/30 | Units |
Film Size | 4” x 4” Std, 2” x 2” optional |
Film Thickness | Up to 120 Mil |
Film Test Area | 50 cm2 standard cartridge |
5.6 cm2 cartridge – optional |
Other Specifications (all models):
Physical Specifications
- 15.5” H x 12” W x 23” D (39.4 cm x 30.4 cm x 58.0 cm)
- 95 Lbs. (43.1 kg)
Electrical Requirements
- Voltage: 100 – 240 VAC 50/60 Hz
- Maximum Power Draw: 700 VA
Laboratory Requirements
Temperature | Operation | 22 °C ± 2 °C |
Storage | 10 °C to 30 °C | |
Humidity | Operation | 20% to 80% RH (non-condensing) |
Storage | 5% to 85% RH (non-condensing) |
Gas Supply Requirements
Carrier Gas | Gas Composition | Nitrogen (99.7% N2 or better) |
Test Gas | Gas Composition | Carbon Dioxide (99.7% or better) |
Connectivity
- Remote Access and Control (requires optional Perm-Net Light software)
- Standard Ethernet port (100T)
- Dual USB 2.0 ports (for connection of keyboard, mouse or memory devices)
Compliance Standards
This product is UL and CSA listed using the ETL mark from Intertek Group plc. and is certified to conform to CE.
Conforms to the following internationally recognized safety standards.
- IEC 61010-1:2010 Third Edition
- UL 61010-1:2012 Third Edition
- CAN/CSA-C22.2 No. 61010-1:2012 Third Edition
Conforms to the following internationally recognized EMC standards.
- EN 61326-1:2013
- EN 61000-3-2:2006, A1:2009, A2:2009
- EN 61000-3-3:2013
Conforms to EU Directive 2011/65/EU RoHS Compliance.
Cartridges & Options
- Reduced area cartridge
- Blister cartridge
- Package testing adaptor
- PackRack package test fixture
- Closure test cell
- 1 Liter bottle capture volume test cell
- 2 Liter bottle capture volume test cell
Consumables
- Reference Films
- Masks
Software
- PermWare data collection, data management and instrument status dashboard
- Remote viewing software from a PC – PermNet Lite
- 21 CFR Part 11 user and data security
Service & Support Options
- IQ/OQ documentation & validation
- Basic Preventative Maintenance Plan
- Plus Preventative Maintenance Plan
Packaging Films
- Monolayer
- Multi-layer
- Coated Films
Packages
- Common food packages
- Bottles/Closures
Test Standard Compliance
- ASTM F2476
- DIN 53380-4
- ISO 15105-2
Test Methodology
Following ASTM F2476-13;
Standard Test Method for the Determination of Carbon Dioxide Gas Transmission Rate (CO2TR) Through Barrier Materials Using an Infrared Detector
Summary of Test Method:
1. The carbon dioxide gas transmission rate is determined after the sample has equilibrated in a dry-test environment. In this context, a “dry” environment is considered to be one in which the relative humidity is less than 1 %.
2. The specimen is mounted as a sealed semi-barrier between two chambers at ambient atmospheric pressure. One chamber is slowly purged by a stream of nitrogen and the other chamber with carbon dioxide. As carbon dioxide gas permeates through the film into the nitrogen carrier gas, it is transported to an infrared detector where an electrical output is produced whose magnitude is proportional to the amount of CO2 flowing into the detector per unit of time.
DATA SHEET