Lippke® 5000
Reliable seal strength and pressure leak testing of many types of packaging.
- Precise, definable and reproducible test conditions
- Suitable for many different kinds of packages
- Burst tester, leak, creep, bubble and combined test options
- Suited for testing in production and laboratory environments
- Extended options for analysis and documentation of test results
Measure seal strength and package integrity with the Dansensor® Lippke® 5000. Perform burst, leak, creep, creep to fail and bubble tests from one instrument – or try the combined test option to conduct a simultaneous leak, creep and burst test on a single package. It is ideal for testing rigid, semi rigid and flexible packaging, as well as porous and laminated foils for the food, pharma and medical industries.
With an extensive array of features, test options and accessories, it meets applicable ISO and ASTM standards. In addition, the instrument features individual logins and different user levels to ensure the data security and traceability required by healthcare and pharmaceutical customers.
The leak test systems provide manufacturers with an idea of how well packages resist pressure. Measuring seal strength and carrying out pressure leak tests are important prerequisites to delivering strong and lasting packaging that have a low risk of bursting under pressure.
Leak detectors:
The Dansensor® Lippke® 5000 is one in a series of leak detectors we offer. Each of the leak detectors in the product family are based on an acknowledged pressure leak and seal strength test method which provide manufacturers with unique opportunities for measuring seal strength and conducting pressure leak tests as part of the Quality Control process.
Features
- Improved measuring algorithm
- Test head with integrated sense probe
- Touchscreen with graphical user interface (GUI)
- Data capturing, storage & export
- Individual user logins and access levels for security
- One-touch documentation print-out with external printer
- USB port for optional label printer (thermal or dot matrix), barcode scanner or keyboard
- Meets ASTM and ISO standards
- IQ / OQ available
- PC software version offers additional features::
- FDA 21 CFR, part 11 compliant
- Printable test reports (PDF)
- Test results shown on graphs
- “Combined test” feature
- Interface with SQL database
Benefits
- Optimized measuring from better pressure regulation
- Shorter filling time
- Improved pre-fill process prevents over-shoot
- Improved testing accuracy and shorter test cycle (new algorithm)
- New design & intuitive touch screen interface
- Improved visual display of test results
- Pump feature to easily inflate flat packages
- Covers 1-5 bar pressure range
- Data storage also available on stand-alone unit
- Easy installation & data sharing
Specifications | |
Package types | Flexible, semi-rigid and rigid packages |
Test types | Burst, leak, creep, bubble, combined test |
Test package size | 1-90 000 ml |
Test time | 1-500 sec |
Measuring range | 10-5000 mbar (0.15-72.5 psi) Low range: 10-350 mbar (0.15-5.0 psi) High range: 350-5000 mbar (5.0-72.5 psi) |
Displayed resolution | 0-1000 mbar: 0.1 mbar 1000-5000 mbar: 1 mbar |
Accuracy of measurement | Low range: ± 0.5 mbar (0.007 psi) or 0.5% of reading High range: ± 5 mbar (0.07 psi) or 0.5% of reading |
Number of test definitions | 999 |
Data collection capacity | Test data from more than 1,000,000 tests |
Data collection type | Exported to CSV file |
Dimensions (WxDxH) and weight | 284 mm x 236.5 mm x 185 mm (11.2″ x 9.3″ x 7.3″) 3.4 kg (7.5 lbs) |
IP class | IP 20 |
Ambient temperature | Operational: +2°C to +35°C Storage: -20°C to +60°C |
Relative humidity | +2°C to +25°C; 10 to 90 %RH (non-condensing) +25°C to +30°C; 10 to 70 %RH (non-condensing) +30°C to +35°C; 10 to 50 %RH (non-condensing) |
Ambient pressure | 750-1100 mbar |
Power supply and consumption | 100-240 VAC, 47-63 Hz Max 40W |
Air supply pressure and connection | 4.0-6.5 bar; at least 1 bar above the test pressure Ø6/4 mm tube DIN ISO 8573-1:2010 [F4:4:3] |
Air consumption | Max 100 L/min (depending on test settings) |
Connectivity | LAN: RJ-45 Ethernet 10/100 Mbit/s, DHCP client or fixed IP USB: 1 x Host, USB 2.0 Type A, 1 x Device, USB 2.0 Type B (max current 500 mA) |
Compliance | CE, China RoHS II |
Standards | ASTM F-1140, F-2054, F-2095, F-2096, ISO 11607 |
Accessories
- Test head, for handheld or stand use
- Choice of blunt or sharp needle
- Stands & Restraining Plates
- Basic stand
- Pneumatic Package Clamp (PPC 300 II) for open packages
- Fixture for ASTM F2054 restraining
- PC software option with CFR 21, part 11 compliance
- IQ/OQ documentation
- Valve Test Unit (VTU) for degassing valves
- Barcode reader (USB)
- Dot matrix printer (USB) for fade-resistant printed records
- Package pressure release valve
- Pressure regulator incl. filter for fine particulates and liquids
- IV-bag adaptor
- Custom-made tube adapters for threaded connections
- Custom-made accessories
Consumables
- Various needles
- Various septa
Applications:
Package integrity tests
- Burst test
- Leak test (pressure decay)
- Creep test (with optional Creep to Fail)
- Bubble test
- Combined test (consecutively on same package)
- Creep and burst
- Creep and leak
- Creep, leak and burst
ASTM Standards:
- ASTM F1140: Burst & creep for medical packages (without restraining plates)
- ASTM F2054: Burst test (with restraining plates)
- ASTM F2095: Pressure decay leak test (with or without restraining plates)
- ASTM F2096: Bubble leak test (with water tank)
Other standards:
- ISO 11607: Packaging for terminally sterilized medical devices
- 21 CFR, part 11: Protects your data and ensures that all records are authentic and incorruptible